B: Troubleshooting
Troubleshooting LR 460 (cont’d)
Symptom | Possible cause | Remedy | |
|
|
| |
|
| Increase TVT manually at vessel bottom | |
Reads empty | Reads vessel bottom | region. | |
when vessel not | through low dK material. | Set Far Range to a lower value. | |
empty. |
| Lower TVT hover level. | |
|
|
| |
Reading | Due to solids’ sloped | Reduce response rate. | |
surfaces some fluctuation | Decrease fill/empty parameter values. | ||
fluctuates. | |||
is normal. | Increase filter time constant in AIFB. | ||
| |||
|
|
| |
Echo has double |
| Increase Echo Reform. | |
peak/is split |
| ||
|
| ||
|
|
| |
Reading is too |
| Increase response rate. | |
| Turn off Window tracking. | ||
slow and lags |
| ||
| Change fill/empty rate parameter | ||
behind actual |
| ||
| values. | ||
level change. |
| ||
| Increase filter time constant in AIFB. | ||
|
| ||
|
|
| |
Device cannot |
| Make sure local operation is enabled. | |
be programmed |
| See Local Operation Enable on page 49. | |
via the |
| Also make sure that Write Locking is | |
handheld |
| Off. See Write Locking on page 49. | |
programmer |
|
| |
|
|
| |
You are unable |
| Make sure remote operation is | |
| enabled. See Remote operation enable | ||
to modify a |
| ||
| (REMLOCK) on page 46. Also make sure | ||
parameter via |
| ||
| that Write Locking is Off. See Write | ||
remote |
| ||
| Locking on page 49. | ||
communications |
| ||
|
| ||
|
|
|
Page | SITRANS LR 460 – INSTRUCTION MANUAL | 7ML19985JM01 |