B: Troubleshooting
General Fault Codes/PROFIBUS PA Extended Diagnostics
Note: Some faults cause the device to go to
LCD |
| Meaning | Corrective Action | Byte | Bit |
display |
| ||||
|
|
|
|
| |
|
| Loss of Echo (LOE) condition: | Ensure installation details |
|
|
|
| the device was unable to get | are correct. Ensure no |
|
|
|
| a measurement within the | material buildup. Adjust |
|
|
S:0 | * | process conditions to |
| 0 | |
|
| Possible causes: faulty | minimize foaming. If problem |
|
|
|
| installation, material buildup, | persists, contact your local |
|
|
|
| and/or presence of foam. | Siemens representative. |
|
|
|
|
|
|
|
|
|
|
| Reset power. If fault persists, |
|
|
S:2 | * | Internal electronics failure. | contact your local Siemens |
| 2 |
|
|
| representative: repair |
|
|
|
|
| required. |
|
|
|
|
|
|
|
|
|
| Device is nearing its lifetime | Replacement is |
|
|
S:3 |
| limit according to the value | recommended. |
| 3 |
| set in Maintenance Required |
| 0 | ||
|
|
|
| ||
|
| Limit. |
|
| |
|
|
|
|
| |
|
|
|
|
|
|
|
| Device is nearing its lifetime | Replacement is |
|
|
S:4 |
| limit according to the value | recommended. |
| 4 |
| set in Maintenance |
|
| ||
|
|
|
|
| |
|
| Demanded Limit. |
|
|
|
|
|
|
|
|
|
|
| Sensor is nearing its lifetime | Replacement is |
|
|
S:6 |
| limit according to the value | recommended. |
| 6 |
| set in Maintenance Required |
|
| ||
|
|
|
|
| |
|
| Limit. |
|
|
|
|
|
|
|
|
|
|
| Sensor is nearing its lifetime | Replacement is |
|
|
S:7 |
| limit according to the value | recommended. |
| 7 |
| set in Maintenance |
|
| ||
|
|
|
|
| |
|
| Demanded Limit. |
|
|
|
|
|
|
|
|
|
Page | SITRANS LR 460 – INSTRUCTION MANUAL | 7ML19985JM01 |