3.3. Sensor Calibration

To perform calibration via SIMATIC PDM open the menu Device – Sensor

Calibration.

Sensor Reference Point1

(flange face)

High Calibration Point

Level2

Low Calibration Point

LR 460 with Easy Aimer

High Level Point (typically 100%)

Low Level Point (typically 0%)

3.3.1. Sensor Units

Whatever units the sensor is measuring in.

Options

m, in, cm, mm, ft

 

default: m

 

 

 

3.3.2. Calibration Type

In Dry calibration, the user enters all four calibration values: High and Low Level Points, and High and Low Calibration Points.

Options

DRY CAL

*

Dry calibration

3.3.3. Low Calibration Pt. (default 30.000 m)

Distance from Sensor Reference to Low Calibration Point (corresponding to Low Level Point). Unit is defined in Sensor units.

1.Sensor Reference Point: the point to which all the Sensor Calibration parameters are referenced, which is the flange face.

2.Level Value: the level measured in level units.

7ML19985JM01