3.3. Sensor Calibration
To perform calibration via SIMATIC PDM open the menu Device – Sensor
Calibration.
Sensor Reference Point1 |
(flange face) |
High Calibration Point |
Level2 |
Low Calibration Point |
LR 460 with Easy Aimer
High Level Point (typically 100%)
Low Level Point (typically 0%)
3.3.1. Sensor Units
Whatever units the sensor is measuring in.
Options | m, in, cm, mm, ft | |
| ||
default: m | ||
| ||
|
|
3.3.2. Calibration Type
In Dry calibration, the user enters all four calibration values: High and Low Level Points, and High and Low Calibration Points.
Options | DRY CAL | * | Dry calibration |
3.3.3. Low Calibration Pt. (default 30.000 m)
Distance from Sensor Reference to Low Calibration Point (corresponding to Low Level Point). Unit is defined in Sensor units.
1.Sensor Reference Point: the point to which all the Sensor Calibration parameters are referenced, which is the flange face.
2.Level Value: the level measured in level units.