Parameters

Quick Start Wizard

The Quick Start wizard groups together all the settings you need to make for a simple application.

The wizard is a complete package and settings are inter-related.

Do not use the wizard to modify individual parameters.

You can access the Quick Start wizard either via SIMATIC PDM, or via the handheld programmer.

There is no need to perform a reset before a Quick Start.

1.Quick Start

In SIMATIC PDM, Open the menu Device – Quick Start, and follow steps 1 to 5.

If you are using the handheld programmer, activate program mode and work through the Quick Start sequence.

1.1. Language

Options

ENGLISH, DEUTSCH, FRANCAIS, ESPANOL

1.2. Application Type

 

 

 

Options

 

STEEL

Silo construction

CONCRETE

 

 

 

1.3. Operation

 

 

 

 

 

LEVEL

 

Distance to material surface referenced from Low

 

 

Calibration Point (process empty level).

 

 

 

Options

SPACE

 

Distance to High Calibration Point (process full level)

 

referenced from material surface.

 

 

 

 

DISTANCE

Distance to material surface referenced from Sensor

 

Reference Point.

 

 

 

Operation types

 

 

 

sensor reference point

 

high

 

(flange face)

 

 

 

 

 

calibration

 

 

 

 

 

 

point

distance

 

 

 

 

 

 

 

 

space

level

low calibration point