Siemens LR 460 High Calibration Pt. default 0.350 m, Unit Level, Low Level Point default

Models: LR 460

1 123
Download 123 pages 8.81 Kb
Page 58
Image 58

Parameters

3.3.4. High Calibration Pt. (default 0.350 m)

Distance from Sensor Reference to High Calibration Point (corresponding to High Level Point). Unit is defined in Sensor units.

3.3.5. Unit (Level)

Selected engineering units for Level (PV), referenced from Low Level Point (plus level offset, if any).

 

percent

%

Options

linear

m, in, cm, mm, ft, in;

 

default: m

 

 

 

 

3.3.6. Low Level Point (default 0.000)

The level when the material is at Low Calibration Point.The unit is defined in Level units.

3.3.7. High Level Point (default 100.000)

The level when the material is at High Calibration Point.The unit is defined in Level units.

3.3.8. Level Offset (default 0.000)

A constant offset that can be added to Level to form PV/SV! (default level output). See page 101 for an illustration. The unit is defined in Level units.

3.3.9. Sensor Offset (default 0)

The offset from the Sensor’s reference point to the vessel’s reference point.This is a constant offset that is subtracted from the Sensor value.The unit is defined in Sensor Units. (See How the Level Transducer Block works: on page 100 for further details.)

Compensates, for example if the sensor head is changed. 3.3.A. Temperature Units

Selects the engineering unit to be displayed with the value representing temperature.

Options

°C, °F, °R, °K

 

default: °C

 

 

 

Page 52

SITRANS LR 460 – INSTRUCTION MANUAL

7ML19985JM01

Page 58
Image 58
Siemens LR 460 High Calibration Pt. default 0.350 m, Unit Level, Low Level Point default, High Level Point default