Siemens LR 460 instruction manual Related parame

Models: LR 460

1 123
Download 123 pages 8.81 Kb
Page 72
Image 72

Parameters

4.3.4.

Min Limit

 

 

 

 

 

 

 

The mA value to correspond with 4.1.7.1. Lower Value .

 

 

 

 

 

 

 

 

 

Values

 

Range:

 

 

 

 

 

 

 

 

 

 

 

 

Default: 3.8 mA

 

 

 

 

 

 

 

 

 

4.3.5.

Max Limit

 

 

 

 

 

 

 

The mA value to correspond with 4.1.7.2. Upper Value .

 

 

 

 

 

 

 

 

 

Values

 

Range:

 

 

 

 

 

 

 

 

 

 

 

 

Default: 20.5 mA

 

 

 

 

 

 

 

 

 

4.3.6.

FS Mode

 

 

 

 

 

 

 

The mA value to be output in the event of a fault or failure.

 

 

 

 

 

 

 

 

 

mA HOLD

*

 

Level remains at last reading.

 

 

 

 

 

 

 

 

Options

mA SET

 

 

User-defined value set in 4.3.7. FS Value

 

 

 

 

 

 

 

 

mA HIGH

 

 

Use Max Limit as material level.

 

 

 

 

 

 

 

 

 

 

 

 

 

mA LOW

 

 

Use Min Limit as material level.

4.3.7.

FS Value

 

 

 

 

 

 

 

A user-defined mA value to be reported when the Fail-safe timer expires.

 

 

 

 

 

 

 

Values

 

Range: 3.5 to 20.5 mA

 

 

 

 

 

 

 

 

 

 

Default: 3.5 mA

 

 

 

 

 

 

 

 

 

4.3.8.

4 mA Trim

 

 

 

 

 

 

 

Calibrates the 4 mA output.

 

 

 

 

 

 

 

 

 

 

 

Values

 

 

Range: 2.0 to 6.0 mA

 

Related parame-

4.3.1. Function

 

 

A mA value derived from the ratio of current level to the range

 

ters

 

 

 

 

 

between Lower Value and Upper Value.

 

 

 

 

 

 

 

 

 

 

 

 

a.Set 4.3.1. Function to MANUAL mode.

b.Set 4.3.3. Man Value to 4 mA.

c.Attach a calibrated meter and check the output at the terminals. Record the remote reading in mA.

d.Enter the recorded value in 4.3.8. 4 mA Trim .

e.Restore 4.3.1. Function to its previous setting.

f.Confirm that the mA output is as expected.

Page 66

SITRANS LR 460 – INSTRUCTION MANUAL

7ML19985JM01

Page 72