Siemens LR 460 Perform service, Has expired, Calibration Point are, Same Different

Models: LR 460

1 123
Download 123 pages 8.81 Kb
Page 87
Image 87

LCD

 

Meaning (cont’d)

Corrective Action

Byte

Bit

display

 

(cont’d)

 

 

Service interval as defined in

Perform service.

 

 

S:8

 

Maintenance Required Limit

 

 

0

 

 

has expired.

 

 

 

 

 

 

 

 

 

 

 

Service interval as defined in

Perform service.

 

 

S:9

 

Maintenance Demanded

 

 

1

 

 

Limit has expired.

 

 

 

 

 

 

 

 

 

 

 

Input parameters High

Check calibration settings of

 

 

 

 

device. Ensure settings for

 

 

 

 

Calibration Point and Low

 

 

S:10

 

High Calibration Point and

 

3

 

Calibration Point are the

 

 

 

Low Calibration Point are

 

 

 

 

same.

 

 

 

 

different.

 

 

 

 

 

 

 

 

 

 

 

 

 

S:11

 

Internal temperature sensor

Return the device to the

 

4

 

failure.

factory.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lower the ambient

1

 

 

 

Internal temperature of the

temperature enough to cool

 

 

 

device has exceeded

the device.

 

 

S:12

*

specifications: it is operating

Fault code will persist until a

 

5

 

 

outside its temperature

manual reset is performed

 

 

 

 

range.

using PDM or the LCD

 

 

 

 

 

interface.

 

 

 

 

 

 

 

 

 

 

 

Check configuration for

 

 

 

 

Upper and lower input values

AIFB1. Ensure that Upper

 

 

S:14

*

(Process Value Scale) for

Value and Lower Value

 

6

 

 

AIFB1 are the same.

(Process Value Scale) are not

 

 

 

 

 

the same.

 

 

 

 

 

 

 

 

 

 

 

Check configuration for

 

 

 

 

Upper and lower input values

AIFB2. Ensure that Upper

 

 

S:15

 

(Process Value Scale) for

Value and Lower Value

 

7

 

 

AIFB2 are the same.

(Process Value Scale) are not

 

 

 

 

 

the same.

 

 

 

 

 

 

 

 

B: Troubleshooting

7ML19985JM01

SITRANS LR 460 – INSTRUCTION MANUAL

Page B-81

Page 87
Image 87
Siemens LR 460 instruction manual Perform service, Has expired, Calibration Point are, Same Different, Device has exceeded