1.4. Units
Select the units for the Quick Start variables (high and low calibration point, and level, distance, or space).
Options | mm, cm, m, in, ft |
1.5. High Calibration Point
Distance from Sensor Reference to High Calibration Point: usually process full level (see illustration to 1.3)
Values | Range: 0.0000 to 100.00 m |
1.6. Low Calibration Point
Distance from Sensor Reference to Low Calibration Point: usually process empty level (see illustration to 1.3)
Values | Range: 0.0000 to 100.00 m |
1.7. Rate
Sets the reaction speed of the device to measurement changes in the target range.
| SLOW | 0.1 m/minute |
Options |
|
|
MED | 1.0 m/minute | |
|
|
|
| FAST | 10.0 m/minute |
|
|
|
Use a setting just faster than the maximum filling or emptying rate (whichever is greater). Slower settings provide higher accuracy: faster settings allow for more level fluctuation
1.8. Apply Changes
In order to save the settings changed in the Quick Start wizard it is necessary to enable Apply Changes.
Options | YES, NO |
Parameters