4.3.9. 20 mA Trim
Calibrates the 20 mA output.
Values | Range: 18 to 24 mA | |
Related parame- | 4.3.1. Function | |
A mA value derived from the ratio of current level to the range | ||
ters | ||
between Lower Value and Upper Value. | ||
| ||
|
|
a.Set 4.3.1. Function to MANUAL mode.
b.Set 4.3.3. Man Value to 20 mA.
c.Attach a calibrated meter and check the output at the terminals. Record the remote reading in mA.
d.Enter the recorded value in 4.3.9. 20 mA Trim .
e.Restore 4.3.1. Function to its previous setting.
f.Confirm that the mA output is as expected.
4.4.Relay Configuration
4.4.1. AIFB
Select AIFB1 or AIFB2. 4.4.2. Function
Handheld programmer Options
DISABLED | * |
MANUAL |
|
LO ALM |
|
HI ALM |
|
ANY ALM |
|
|
4.4.3. NC/NO
| Handheld programmer | NO | * | Normally Open |
| options | NC |
| Normally Closed |
|
|
|
|
|
4.4.4. State |
|
|
| |
|
|
|
|
|
| Handheld programmer | OFF | * | Relay |
| options | ON |
| Relay energized |
|
|
|
|
|