C: Maintenance
Appendix C: Maintenance
SITRANS LR 460 requires no maintenance or cleaning under normal operating conditions.
Under severe operating conditions, the antenna may require periodic cleaning. If cleaning becomes necessary:
•We recommend using high pressure air to clean the antenna.
•The inside of the horn can be scraped, but be very careful not to damage the small PTFE emitter inside the horn.
Unit Repair and Excluded Liability
All changes and repairs must be done by qualified personnel, and applicable safety regulations must be followed. Please note the following:
•The user is responsible for all changes and repairs made to the device.
•All new components must be provided by Siemens Milltronics Process Instruments Inc.
•Restrict repair to faulty components only.
•Do not
Page | SITRANS LR 460 – INSTRUCTION MANUAL | 7ML19985JM01 |