8.4 Fixed Value Output (Loop Check)

The fixed value output is the function to output a fixed current and/or a fixed pulse output independently of the flow rate signal.

The following parameters can be set here:

(1)Fixed value output On/Off setting

(2)Fixed current output value

(3)Fixed pulse output value

Notes

Fixed pulse output can be sent out only when DO1 or DO2 is set for pulse output.

When the mode is switched to the fixed output mode by DI function, the preset values of items (2) and (3) above will be output.

In the fixed output mode, the condition of DO output (other than pulse output) will be maintained.

Fixed value output On/Off setting

Fixed output mode can be selected by the setting as follows.

Selection

Mode

 

 

1: ON

Fixed value output mode On

 

Current output and pulse output become fixed output values.

 

 

2: OFF

Fixed value output mode Off

 

Fixed output mode will be turned off and the mode goes to the

 

normal measurement mode.

 

 

When the fixed value output is set to ON, the main indicating unit only is displayed and the current output set value is displayed in the lower line in the measurement mode.

Display example:

1.000 m/s * FIX.OUT 20.0 mA

Flow velocity is displayed in main indicating unit

Fixed current value is displayed

Note: When power is turned off and on again, the fixed output On/Off setting will be turned off.

Fixed current output value

The current value used in the fixed output mode is set.

Fixed current output value can be set in the range of 3 to 24mA in increments of 0.1mA.