10. Function Description
The LF232 flowmeter converter is equipped with 4 digital outputs and 2 digital inputs (1 digital output as standard), enabling you to use these as various functions such as pulse output and alarm outputs.
Digital I/O functions are described below.
Functions | Description |
|
|
Flow totalization | Totalizes flow volume in volumetric unit. |
| For totalized flow, a pulse output per each unit of flow volume can be |
| sent out. |
| The operation to start, stop and reset the totalizer and pulse output can |
| be controlled with Digital Input. |
|
|
Multiple ranges | Measuring ranges can be switched in accordance with flow rate. |
| Measuring ranges can be switched automatically or by control using |
| Digital Inputs. |
|
|
Forward and reverse ranges | Forward and reverse direction flows can be measured. |
| Forward and reverse ranges can be combined with multiple ranges. |
|
|
Flow rate High/Low limit | If the flow rate exceeds or lowers below the preset value, an alarm will |
be output. | |
alarms | 2 high limit alarm outputs and 2 low limit alarm outputs, 4 alarm |
| outputs in total, can be sent out. |
|
|
Fluid empty alarm | If the fluid to be measured flows out from the measuring pipe of the |
| detector, an alarm is output. |
| (In the case of LF232*F, fluid empty alarm function is not usable.) |
|
|
Preset counter | If the count of the totalizer exceeds the preset value, a signal is output. |
|
|
Remote still water zero adjustment | Still water zero adjustment can be controlled using Digital Inputs. |
|
|
Fixed value output (loop output) | Fixed value for current output and pulse output can be sent out to |
| perform a loop check for output lines. |
| The fixed value output mode can also be switched by control using |
| Digital Inputs. |
|
|
Converter error alarm | If an error is detected such as memory error or excitation circuit error, |
| an alarm is output. |
|
|