9.2.3100% Flow Rate Calibration (Span Calibration)

Using the Mag-Prover internal calibration circuit, 100% flow rate value (hereafter called the span value) calibration can be performed.

Checking the span value

Switch operation

Display example

Description

 

 

 

 

P3:FLOW CAL100

Select P3:FLOW CAL 100 from the calibration item selection

 

menu.

 

100.1 %

 

Span value using the simulation input appears.

 

 

P3:FLOW CAL100

Returns to the menu of calibration item selection.

Calibrating the span value

Switch operation

Display example

 

Description

 

 

 

 

P3:FLOW CAL100%

Select P3:FLOW CAL 100 from the calibration item selection

 

menu.

 

 

100.1 %

 

 

Span value using the simulation input appears.

 

 

 

 

 

 

 

ADJUST READY

Press

. Then the display shown left appears and the system

 

100.1 %

is ready for span calibration. (Note 1)

 

 

 

 

 

P3:FLOW CAL100

The display shown left appears and the span calibration starts.

 

* CAL. 100% ADJ.

 

 

 

 

 

 

P3:FLOW SCAL100

About 3 to 6 seconds later (Note 2), the span calibration ends

100.0 %

and a new span value appears.

 

P3:FLOW CAL100

Returns to the menu of calibration item selection.

Note 1: To cancel the adjustment from the condition of ADJUST READY, press CANCEL . The display returns to the condition where 100% flow rate value using the simulation input is displayed.

Note 2: Calibration time differs depending on the exciting frequency.