8.5.1 Still Water Zero Adjustment
Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector
stays still.
Switch operation Display example Description
−
10.00 m/s
100.0 % Measured value being displayed (measurement mode)
ENTER111
000
Pressing MENU/ENT in the measurement mode for 3 seconds
or more, the screen moves to the menu lock cancellation
screen.
* Pressing CANCEL in the mode selection screen, the mode
returns to the measurement mode.
ENTER111
111
Set the menu lock cancellation cord [111] by
pressing and .
* Pressing CANCEL in the mode selection screen, the mode
returns to the measurement mode.
>1:SET
2:CNT CTRL
Press MENU/ENT for 3 seconds or more in the measurement
mode and then the mode changes to the mode selection screen.
A1:EX CURRENT Select 1:SET in the mode selection screen and then the menu
screen appears.
E1: ZERO ADJUST Press and so that the function number becomes E1:
ZERO ADJUST.
E1: ZERO ADJUST
01.0%
The current flow rate measured value in % unit appears.
(Check that the fluid to be measured stays still.)
ADJUST READY
01.0%
ADJUST READY appears and the system is ready for
adjustment. (If “CANCEL” is pressed, Adjustment ready
condition will be cancelled.)
E1: ZERO ADJUST
* ZERO ADJUST
About 3 to 6 seconds later, the still water zero adjustment ends
and the measured flow value after adjustment appears.
−
E1: ZERO ADJUST
00.0% Returns to the menu of setting item selection.
E1: ZERO ADJUST Press and so that the number digit of the function
number becomes 0.