8.5 Zero Adjustment
8.5.1Still Water Zero Adjustment
Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector stays still.
Note2
Note2
Switch operation | Display example |
|
|
|
|
|
|
| Description | |
|
|
|
|
|
|
|
|
|
| |
− | 10.00 m/s | Measured value being displayed (measurement mode) | ||||||||
100.0 % | ||||||||||
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| ||
|
| Pressing |
|
|
|
|
| in the measurement mode for 3 seconds | ||
|
| MENU/ENT | ||||||||
| ENTER111 | or more, the screen moves to the menu lock cancellation | ||||||||
| screen. |
|
|
|
|
|
| |||
| 000 |
|
|
|
|
|
| |||
| * Pressing | CANCEL | in the mode selection screen, the mode | |||||||
|
| |||||||||
|
| returns to the measurement mode. | ||||||||
|
|
|
|
| ||||||
|
| Set the menu lock cancellation cord [111] by | ||||||||
| ENTER111 | pressing |
| and | . |
|
| |||
| 111 | * Pressing |
|
| in the mode selection screen, the mode | |||||
| CANCEL | |||||||||
|
| returns to the measurement mode. | ||||||||
|
|
|
|
| ||||||
|
| Press |
|
|
|
| for 3 seconds or more in the measurement | |||
| >1:SET | MENU/ENT | ||||||||
| 2:CNT CTRL | mode and then the mode changes to the mode selection screen. | ||||||||
|
|
| ||||||||
| A1:EX CURRENT | Select 1:SET in the mode selection screen and then the menu | ||||||||
| screen appears. |
|
|
|
| |||||
|
|
|
|
|
| |||||
|
|
|
|
| ||||||
| E1: ZERO ADJUST | Press | and | so that the function number becomes E1: | ||||||
| ZERO ADJUST. |
|
|
|
| |||||
|
|
|
|
|
| |||||
|
|
| ||||||||
| E1: ZERO ADJUST | The current flow rate measured value in % unit appears. | ||||||||
| 01.0% | (Check that the fluid to be measured stays still.) | ||||||||
|
|
| ||||||||
| ADJUST READY | ADJUST READY appears and the system is ready for | ||||||||
| adjustment. (If “CANCEL” is pressed, Adjustment ready | |||||||||
| 01.0% | |||||||||
| condition will be cancelled.) | |||||||||
|
| |||||||||
|
|
| ||||||||
| E1: ZERO ADJUST | About 3 to 6 seconds later, the still water zero adjustment ends | ||||||||
| * ZERO ADJUST | and the measured flow value after adjustment appears. | ||||||||
|
|
|
|
|
|
|
|
|
| |
− | E1: ZERO ADJUST | Returns to the menu of setting item selection. | ||||||||
00.0% | ||||||||||
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| ||||||
| E1: ZERO ADJUST | Press | and | so that the number digit of the function | ||||||
| number becomes 0. | |||||||||
|
| |||||||||
|
|
|
|
|
|
|
|
|
|