6F8A0917
12.2Output When an Error or Alarm Occurs
Error indication | Measured value | Current output | Totalizer pulse | Remarks | |
display | (4 to 20mA) | output | |||
|
| ||||
|
|
|
|
| |
ROM ERROR (Note 1) | − | 4mA (standard) | Stopped | After | |
(Note 3) |
| does not start. | |||
|
|
| |||
|
|
|
|
| |
RAM ERROR | − | 4mA (standard) | Stopped | After | |
(Note 3) |
| does not start. | |||
|
|
| |||
|
|
|
|
| |
PARAMETER FAIL | Zero | 4mA (standard) | Stopped | − | |
(Note 2) |
| (Note 3) |
| ||
|
|
| |||
|
|
|
|
| |
EX.CURR OPEN | Zero | 4mA (standard) | Stopped | Still water zero adjustment | |
| (Note 3) |
| cannot be conducted. | ||
|
|
| |||
|
|
|
|
| |
EX.CURR ERROR | Zero | 4mA (standard) | Stopped | − | |
| (Note 3) |
| |||
|
|
|
| ||
|
|
|
|
| |
ADC ERROR | Zero | 4mA (standard) | Stopped | − | |
| (Note 3) |
| |||
|
|
|
| ||
|
|
|
|
| |
EMPTY | Zero | 4mA (standard) | Stopped | Still water zero adjustment | |
| (Note 3) |
| cannot be performed. | ||
|
|
| |||
|
|
|
|
| |
INVALID TOTAL | Measured value | Measured value | Measured value | The error message disappears if | |
|
|
| you clear (reset) the totalizer. | ||
|
|
|
| ||
|
|
|
|
| |
HIGH ALARM | Measured value | Measured value | Measured value | − | |
|
|
|
|
| |
Measured value | Measured value | Measured value | − | ||
|
|
|
|
| |
L ALARM | Measured value | Measured value | Measured value | − | |
|
|
|
|
| |
Measured value | Measured value | Measured value | − | ||
|
|
|
|
|
Note 1: The display and output may be uncertain depending on the nature of the ROM error.
Note 2: If a parameter failure relating to the current output occurs, the current output may not become exactly the set value of the current output used when an alarm occurs.
Note 3: The output becomes the value selected in 8.2.11 “Current Output Setting Used When an Alarm Occurs.”
Note 4: If the fluid empty alarm not included (WITHOUT EMP) is selected (standard) in the setting for alarm output factors in 8.2.19