12.2 Output When an Error or Alarm Occurs
Error indication Measured value
display
Current output
(4 to 20mA)
Totalizer pulse
output Remarks
ROM ERROR (Note 1)
4mA (standard)
(Note 3)
Stopped After power-up, measurement
does not start.
RAM ERROR
4mA (standard)
(Note 3)
Stopped After power-up, measurement
does not start.
PARAMETER FAIL
(Note 2)
Zero 4mA (standard)
(Note 3)
Stopped
EX.CURR OPEN Zero 4mA (standard)
(Note 3)
Stopped Still water zero adjustment
cannot be conducted.
EX.CURR ERROR Zero 4mA (standard)
(Note 3)
Stopped
ADC ERROR Zero 4mA (standard)
(Note 3)
Stopped
EMPTY Zero 4mA (standard)
(Note 3)
Stopped Still water zero adjustment
cannot be performed.
INVALID TOTAL Measured value Measured value Measured value The error message disappears if
you clear (reset) the totalizer.
HIGH ALARM Measured value Measured value Measured value
H-HIGH ALARM Measured value Measured value Measured value
L ALARM Measured value Measured value Measured value
L-LOW ALARM Measured value Measured value Measured value
Note 1: The display and output may be uncertain depending on the nature of the ROM error.
Note 2: If a parameter failure relating to the current output occurs, the current output may not become exactly
the set value of the current output used when an alarm occurs.
Note 3: The output becomes the value selected in 8.2.11 “Current Output Setting Used When an Alarm
Occurs.”
Note 4: If the fluid empty alarm not included (WITHOUT EMP) is selected (standard) in the setting for alarm
output factors in 8.2.19 “Self-Diagnosis Function,” the output becomes 4mA and if the fluid alarm
included (WITH EMP) is selected, the output becomes the same as the above in Note 3.