6F8A0917
8.2.6Normal Indicating Unit
You can select one of the units listed below as normal indicating unit for 2 types of units, main unit (UNIT 1) and sub unit (UNIT 2) to be displayed in the measurement mode.
•Flow velocity unit: m/s, ft/s
• Flow rate unit: | m3/s, m3/min, m3/h, m3/d |
| l/s, l/min, l/h, l/d |
| ml/s, ml/min, ml/h, ml/d |
| gal/s, gal/min, gal/h, gal/d |
| bbl/s, bbl/min, bbl/h, bbl/d |
| pt/s, pt/min, pt/h, pt/d |
| qt/s, qt/min, qt/h, qt/d |
•Volumetric flow: m3, l, ml, gal, bbl, pt, qt
• Other units: | %, COUNT, RANGE, CUSTOM |
%: | Indicates the value in percent of the setting range |
| When |
COUNT: Indicates the count of the totalizer.
RANGE: Indicates the range number of the range in operation when
CUSTOM: Indicates the flow rate value in the unit defined in 8.2.7 “Custom Unit.”
If the unit is set to volumetric flow or COUNT, the flow direction for total flow can be set.
If the indicting unit is set to volumetric flow or COUNT, the flow direction for total flow can also be set.
• Volumetric flow direction: F: Forward direction, R: Reverse direction
B:Forward/Reverse automatic switching, D: Total difference flow indication
Checking the normal indicating unit
Proceed as follows to check the normal indicating unit.
Here, only about the main unit (UNIT1) is explained. For sub unit (UNIT2), select the function number “B2”.
Switch operation | Display example | Description |
|
|
|
| B1: UNIT 1 | Select B1: UNIT 1 from the setting item selection menu. |
%The currently set main indicating unit appears.
B1: UNIT 1
Returns to the menu of setting item selection.
- 66 -