Changing the fixed output function

The following is an example to change the fixed output function from Fixed value output OFF to ON and then go on to set the current output value and the pulse output value. To change the current output value or pulse output value independently, select the following function:

Fixed current output value

 

N2: FIX CURR

 

Fixed pulse output value

 

N3: FIX PULSE

 

 

 

 

 

 

Switch operation

Display example

 

 

Description

 

 

 

 

 

 

N1: FIXED OUT

Select N1: FIXED OUT from the setting item selection menu.

 

2:OFF

The currently set fixed value output On/Off setting appears.

 

 

 

 

 

 

 

N1: FIXED OUT

The cursor appears.

 

 

2:OFF

 

 

 

 

 

 

 

 

 

 

 

 

 

N1: FIXED OUT

Select the fixed value output: ON.

 

1:ON

 

 

 

 

 

 

 

 

 

 

The display blinks and the set value is displayed for

 

N2: FIX. CURR.

conformation.

 

 

Press

MENU/ENT

again to write the data. The cursor

 

10.0 mA

 

disappears and then the current output set value can be

 

 

 

 

changed.

 

 

N2: FIX. CURR.

Move the cursor to the desired digit to change.

 

10.0 mA

 

 

 

 

 

 

 

 

 

N2: FIX. CURR.

Change the number of the digit.

 

To change other digit, press

under this condition to move

 

20.0 mA

 

the cursor to the desired digit and change the number there.

 

 

 

 

 

 

 

The display blinks and the set value is displayed for

 

N3: FIX. PULSE

conformation.

 

 

100 PPS

Press

MENU/ENT

again to write the data. The cursor

 

 

disappears and then the pulse output set value can be changed.

 

 

 

 

 

The display blinks and the set value is displayed for

 

N3: FIX. PULSE

conformation.

 

 

Press

MENU/ENT

again to write the data. Then the cursor

 

100 PPS

 

disappears and the display returns to the condition that the

 

 

 

 

setting value is displayed.

 

N3: FIX. PULSE

Returns to the menu of setting item selection.

Note. If a value outside of the settable range is tried to set, the upper limit or lower limit value of the settable range will be forcibly set.