5.4.3Grounding
The external ground terminal of the detector and the FG terminal of the converter (or external ground terminal of the converter) must be grounded securely with Class D grounding (grounding resistance 100Ω or less). Use an IV wire 5.5mm2 or more for grounding wire.
In addition, do not share the grounding wire with other equipment where grounding current may flow. (An independent grounding is preferable.)
If it is difficult to carry out grounding work at the detector side due to a pit installation or other reasons, use a
Converter | A | B | X |
|
terminal block | G | E | Y | FG |
External ground |
|
|
|
|
terminal |
|
|
|
|
Twisting shields |
|
|
| Class D grounding |
|
|
|
| |
|
|
|
| (Ground resistance 100 or less) |
Input signal cable |
|
|
|
|
Twisting shields |
|
|
| Excitation cable |
|
|
|
| Ground terminal |
|
| G |
| E |
| A | B | X | Y |
|
|
|
| Class D ground |
|
|
|
| (Ground resistance 100 or less) |
Detector terminal block
Figure 5.9(a) Wiring between Detector and Converter