6F8A0917
-
57
-
Changing the exciting current value
To change the exciting current value, proceed as follows. However, the exciting current value is already
adjusted when the product is shipped from the factory. Do not change the exciting current value unless the
value is different from the one described on the detector nameplate. If you change this value to other value,
this may cause an error.
The following is an example to change the exciting current value from 0.1900A to 0.2150A.
Switch operation Display example Description
A1: EX. CURR.
0.1900A
Select A1:EX. CURR. from the setting item selection menu.
The currently set exciting current value (0.1900A in this
example) appears.
A1: EX. CURR.
0.1900A
The cursor appears.
Press further to move the cursor to the desired digit.
A1: EX. CURR.
0.2900A
0.2100A
0.2150A
Change the number of the digit.
To change the number of other digit, press to move the
cursor to that digit and change the number there. Repeat this
operation until 0.2150A appears on the display.
A1: EX. CURR.
0.2150A
The display blinks and the set value is displayed for
confirmation.
Press MENU/ENT again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
A1: EX. CURR. Returns to the menu of setting item selection.
Note: Set the exciting current value within the range shown below:
For small and medium meter size detectors (LF232*A) 0.0500A to 0.2300A
For large meter size detectors (LF232*B) 0.0500A to 3.000A
For partially filled pipes detectors (LF232*F) 0.0500A to 1.500A
If you set an exciting current value outside of the settable range, a message such as *H. OVER
SPEC. appears and the value returns to the previous value. Try to set a new value again.