6F8A0917
13.2.2 Flow rate indication is not correct
START
Is the velocity range correctly set?
YES
Is zero point correctly set?
YES
NO
NO
Set it correctly according to 8. “Parameter Settings/Adjustment.”
Perform zero adjustment according to 6. “Operation” and 8.5 “Zero Adjustment.”
Is the exciting current correct?
YES
NO
Set the exciting current correctly according to 8. “Parameter Settings.”
Isn't the inside wall of detector pipe or electrode contaminated?
NO
Is the output load resistance 1k or less?
YES
Aren't there two load resistors connected to the output in parallel?
NO
Is accuracy calculated as follows?
(Measured flow rate) (Actual flow rate)
Actual flow rate
100 (%)
YES
Contact your nearest Toshiba representative.
YES
NO
YES
NO
Clean the inside wall of the detector pipe to remove the deposits.
Design the output circuit so that the output load resistance is less than 1 k.
Connect the two load resistors in series, if necessary.
Calculate as shown on the left.