6F8A0917
-
121
-
Changing the zero offset adjustment value
Changing the zero offset adjustment value from +1.0% to 2.5%.
Switch operation Display example Description
O1: MANUAL ZERO
+001.0%
Select O1: MANUAL ZERO from the setting item selection menu.
The currently set zero offset adjustment value appears.
O1: MANUAL ZERO
+001.0%
The cursor appears.
Press further to change the cursor to the desired digit to
change.
O1: MANUAL ZERO
-001.0%
-002.5%
Change the number of the digit.
To change other digit, press under this condition to move
the cursor to that digit and change the number there.
O1: MANUAL ZERO
-002.5%
The display blinks and the set value is displayed for
confirmation.
Press MENU/ENT again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
O1: MANUAL ZERO Returns to the menu of setting item selection.
Note: If you try to set a value outside of the settable range, an error message of *H. OVER SPEC or *L. OVER
SPEC appears. Try to set a new value again.