9.2 Converter Check / Calibration

9.2.10% Flow Rate Calibration (Zero Calibration)

Using Mag-Prover internal calibration circuit, 0% flow rate (hereafter called zero point) calibration can be performed.

Checking the zero point

Switch operation

Display example

Description

 

 

 

 

 

P1:FLOW CAL

0

Select P1:FLOW CAL 0 from the calibration item selection

 

menu.

 

0.0

%

 

Zero point using the simulation input appears.

 

 

 

P1:FLOW CAL 0

Returns to the menu of calibration item selection.

Calibrating the zero point

Switch operation

Display example

 

Description

 

 

 

 

 

 

P1:FLOW

CAL

0

Select P1:FLOW CAL 0 from the calibration item selection

 

menu.

 

 

 

0.1

%

 

 

 

Zero point using the simulation input appears.

 

 

 

 

 

 

 

 

 

 

ADJUST READY

 

Press

. Then the display shown left appears and the system

 

 

0.1

%

is ready for zero calibration. (Note 1)

 

 

 

 

 

 

 

P1:FLOW

CAL

0%

The display shown left appears and zero calibration starts.

 

* CAL.

0% ADJ.

 

 

 

 

 

 

 

 

P1:FLOW

CAL

0

About 3 to 6 seconds later (Note 2), the zero calibration ends

 

0.0 %

and a new zero point appears.

 

 

P1:FLOW CAL 0%

Returns to the menu of calibration item selection.

Note 1: To cancel the adjustment from the condition of ADJUST READY, press CANCEL to the condition where zero point using the simulation input is displayed.

Note 2: Calibration time differs depending on the exciting frequency.

. The display returns