9.2 Converter Check / Calibration
9.2.10% Flow Rate Calibration (Zero Calibration)
Using
Checking the zero point
Switch operation | Display example | Description | |
|
|
|
|
| P1:FLOW CAL | 0 | Select P1:FLOW CAL 0 from the calibration item selection |
| menu. | ||
| 0.0 | % | |
| Zero point using the simulation input appears. | ||
|
|
|
P1:FLOW CAL 0
Returns to the menu of calibration item selection.
Calibrating the zero point
Switch operation | Display example |
| Description | ||
|
|
|
|
| |
| P1:FLOW | CAL | 0 | Select P1:FLOW CAL 0 from the calibration item selection | |
| menu. |
| |||
|
| 0.1 | % |
| |
|
| Zero point using the simulation input appears. | |||
|
|
|
| ||
|
|
|
|
| |
| ADJUST READY |
| Press | . Then the display shown left appears and the system | |
|
| 0.1 | % | is ready for zero calibration. (Note 1) | |
|
|
|
|
|
|
| P1:FLOW | CAL | 0% | The display shown left appears and zero calibration starts. | |
| * CAL. | 0% ADJ. | |||
|
|
| |||
|
|
|
|
| |
− | P1:FLOW | CAL | 0 | About 3 to 6 seconds later (Note 2), the zero calibration ends | |
| 0.0 % | and a new zero point appears. | |||
|
|
P1:FLOW CAL 0%
Returns to the menu of calibration item selection.
Note 1: To cancel the adjustment from the condition of ADJUST READY, press CANCEL to the condition where zero point using the simulation input is displayed.
Note 2: Calibration time differs depending on the exciting frequency.
. The display returns