6F8A0917

Pulse width

The pulse width for totalizer pulse output is set.

If Automatic setting (1:AUTO) is selected for pulse width mode, it is not necessary to set the pulse width.

The pulse width must be set to a value within the range of 0.3ms to 500ms. If you try to set a value exceeding 500 ms, the value will be forcibly changed to 500ms.

The pulse width must be set to 40% or less of the pulse period at 100% output. If a value exceeding this limit is tried to set, regardless of the range described above, an error message

*H. OVER SPEC

appears and the value returns to the previously set value. Try to set a new value again.

If the pulse width is set to 0, it will be automatically set to 40% of the pulse period at 100% output. However, if the calculation result exceeds 100ms, it will be forcibly set to 100ms.

Example 1 In the case of the following: Range: 3,600m3/h (1m3/s) Count rate: 0.001m3

Since the count rate is 3,600 m3/h ÷ 0.001(m3) = 3,600,000 pulses/h (1000 pulses/s), the full-scale period is 1s ÷ 1000 = 1ms.

Therefore, the pulse width can only be set to 1ms x 0.4 = 0.4ms.

Example 2 In the case of the following: Rage: 3,600m3/h (1m3/s) Count rate: 1000m3

Since the pulse count is 3,600m3/h ÷ 1000m3 = 3.6 pulses/h (0.001 pulse/s), the full-scale period is 1s ÷ 0.001 = 1000s.

Therefore, the pulse width becomes 1000s × 0.4 = 400s. However, since the maximum value is 500ms, the pulse width becomes 500ms.

Example 3 In the case of the following: Rage: 3,600m3/h (1m3/s) Count rate: 1m3

Pulse width is set to 0ms

Since the pulse count is 3,600m3/h ÷ 1m3 = 3600 pulses/h (1 pulse/s), the full-scale period is 1s ÷ 1 = 1s.

Therefore, the pulse width becomes 1s × 0.4 = 400ms. However, in the case of automatic setting, since the maximum value is 100ms, the pulse width becomes 100ms.

- 94 -

Page 95
Image 95
Toshiba L5232 manual Pulse width, Over Spec