6F8A0917
-
95
-
Checking the count rate and pulse width
How to check the count rate setting is shown below.
To check other setting values, select the following function numbers:
Pulse width mode H2: PLS MODE
Pulse width H3: PLS WIDTH
Switch operation Display example Description
H1: COUNT RATE
1.00000 m3
Select H1: COUNT RATE from the setting item selection
menu.
The currently set count rate appears.
H1: COUNT RATE Returns to the menu of setting item selection.
Changing the count rate
The following is an example to change the count rate from 0.01m3 to 200 l (liter).
Switch operation Display example Description
H1: COUNT RATE
1.00000 m3
Select H1: COUNT RATE from the setting item selection
menu.
The currently set count rate appears.
H1: COUNT RATE
1.00000 m3
The cursor appears.
Press further to move the cursor to the digit of unit.
H1: COUNT RATE
1000.00 l
200.000 l
Change the unit to l.
Then press to move the cursor to the desired digit to change
and change the number there. Repeat this operation until 200 l
appears on the display.
H1: COUNT RATE
200.000 l
The display blinks and the set value is displayed for
confirmation.
Press MENU/ENT again to write the data. Then the cursor
disappears and the display returns to the condition that the set
value is displayed.
H1: COUNT RATE Returns to the menu of setting item selection.
Note: The indicting units for count rate are displayed cyclically as shown below.
m
3 ml
l gal bbl pt qt