6F8A0917

7.4 Setting Menu List

The constants set in the LF232 converter can be checked or changed in the order shown in the table below.

The details of each item are described in the following sections:

Setting items (A to M): Chapter 8. “Parameter Settings / Adjustment”

Calibration item (N): Chapter 9. “Mag-Prover Calibration”

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

P

0

Return to measurement mode

1

2

3

4

5

6

7

8

 

 

 

 

 

 

 

 

Exciting

Meter size *1

Exciting

Flow direction

Password *1

 

 

 

current setting

 

frequency *1

setting *1

 

 

 

 

*1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Main display

Sub display

Custom

Custom unit

 

 

 

 

unit

unit

coefficient *1

*1

 

 

 

 

 

 

 

 

 

 

 

 

Range type *1

Range 1 *1

Range 2 *1

Range 3 *1

Range 4 *1

Hysteresis *1

 

 

 

 

 

 

 

 

 

 

Damping

Low cutoff

Current output

Display low

Output low

 

 

 

constant

value

setting for

cut

limit setting

 

 

 

 

 

alarm *1

On/Off

*1

 

 

 

 

 

 

 

 

 

 

 

Still water

 

 

 

 

 

 

 

zero

 

 

 

 

 

 

 

adjustment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Digital output

Digital output

Digital output

Digital output

DO1 alarm

DO2 alarm

DO3 alarm

DO4 alarm

1 *1

2 *1

3 *1

4 *1

output state *1

output state *1

output state *1

output state *1

 

 

 

 

 

 

 

 

Digital input 1

Digital input 2

DI1 control

DI2 control

 

 

 

 

*1

*1

signal level

signal level

 

 

 

 

 

 

setting *1

setting *1

 

 

 

 

 

 

 

 

 

 

 

 

Count rate *1

Pulse width

Pulse width *1

 

 

 

 

 

 

setting mode

 

 

 

 

 

 

 

*1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Preset count

Preset output

 

 

 

 

 

 

value *1

setting *1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

High limit

High limit

Low limit

Low limit

 

 

 

 

alarm On/Off

value setting

alarm On/Off

value setting

 

 

 

 

*1

*1

*1

*1

 

 

 

 

 

 

 

 

 

 

 

 

High high

High high

Low low limit

Low low limit

 

 

 

 

limit alarm

limit value

value On/Off

value setting

 

 

 

 

On/Off *1

setting *1

*1

*1

 

 

 

 

 

 

 

 

 

 

 

 

Fluid empty

Self-diagnosis

Alarm output

 

 

 

 

 

alarm *1

On/Off *1

preset *1

 

 

 

 

 

 

 

 

 

 

 

 

 

Rate-of-chang

Control time

 

 

 

 

 

 

e limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fixed value

Fixed current

Fixed pulse *1

 

 

 

 

 

output *1

*1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero offset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0%

50%

100%

Exciting

 

 

 

 

Flow value

Flow value

Flow value

current

 

 

 

 

calibration *1

calibration

calibration *1

monitor

 

 

 

 

 

 

 

 

 

 

 

 

Note 1: For items marked by *1 in the table above, if you enter a wrong password, you can check the setting value but you cannot change or adjust the setting value.

- 54 -

Page 55
Image 55
Toshiba L5232 manual Setting Menu List, Return to measurement mode