6F8A0917
13. Maintenance and Inspection
Do not conduct wiring work when power is applied.
Wiring while power is applied can cause electric shock.
DON’T
Do not touch the LF232 main body when high temperature fluid is being measured.
| The fluid raises the main body |
| temperature and can cause burns |
DON’T | when touched. |
|
13.1Maintenance
Mag-Prover Calibration / Check
The LF232 converter has a
Check / Replacement of fuse
Fuse can be taken out by turning the top of the fuse holder counterclockwise. Check that the fuse is not damaged. In addition, fuse has its own life and it must to be replaced periodically. (Recommended replacement cycle is about 3 years.)
Type of fuse used: Glass tube fuse (normal blow type)
Rating: | In the case of power supply 100 to 240 VAC and 110VDC | |
| 1A, 250V | 2 pieces |
| In the case of power supply 24VDC and 100 to 120 VAC for large meter size | |
| detector |
|
| 2A, 250V | 2 pieces |
| In the case of power supply 24VDC for large meter size detector and 24VDC | |
| for partially filled pipes detector | |
| 3A, 250V | 2 pieces |
Dimensions: | Diameter 5.2 mm × 20 mm |
Check / Replacement of display unit
If the characters displayed on the LCD become thin or blots come out, the life of the LCD is reached. Replace the display unit with a new one. In order to use the display unit stably for a long time, it is preferable to replace it early. For inspection and replacement, please contact your nearest Toshiba representative.
- 157 -