6F8A0917
8.2.14 Digital I/O Function
Digital I/O functions can be selected.
Select the digital I/O functions from the items shown below.
For details of digital I/O functions, see 10. “Function Description.”
Digital output functions (corresponding to each terminal)
DO FUNCT. | Digital output functions | DO1 | DO2 | DO3 to DO4 |
1: NO USE | Not used |
|
|
|
2: HIGH ALM | High limit alarm output |
|
|
|
3: HH ALM |
|
|
| |
4: LOW ALM | Low limit alarm output |
|
|
|
5: LL ALM |
|
|
| |
6: RNG SIG 1 | Range output No.1 |
|
|
|
7: RNG SIG 2 | Range output No.2 |
|
|
|
8: PRESET | Preset counter output |
|
|
|
9: CONV. ALM | Converter error alarm output (Note 3) |
|
|
|
A: EMPTY ALM | Fluid empty alarm output |
|
|
|
B: PULSE OUT | Pulse output (Note 1) |
| (Note 2) | − |
C: PULSE FRD | Forward flow pulse output |
| (Note 2) | − |
D: PULSE REV | Reverse flow pulse output |
| (Note 2) | − |
|
|
| ○: Enabled | − : Disabled |
Note 1: When the range type is set to the forward/reverse
Note 2: Pulse output using the DO2 can be sent out up to 100pps.
When the pulse output is sent out using the DO2, set the count rate so that the pulse output for full scale value does not exceed 100pps.
For count rate setting, see 8.2.15 “Count Rate (Pulse Rate) and Pulse Width.”
Note 3: For the types of alarms to output, see the alarm output factor setting in 8.2.19
Digital output state (only when alarm output is set)
For alarm output function (items indicated byabove), alarm output operation can be selected independently for each DO terminal.
DO ALM ACT | Alarm output operation |
|
|
1: NORMAL CLOSE | Normal: Contact closed, When alarm occurs: Contact open |
|
|
2: NORMAL OPEN | Normal: Contact open, When alarm occur: Contact closed |
|
|
- 89 -