6F8A0917

No.

Setting item

Display example

 

 

 

 

 

8.2.16

Preset count value

I1: PRESET CNT

00009000

Output function setting

I2: PRESET FNC

1:HOLD

 

 

 

 

 

 

 

 

 

High/Low limit alarm On/Off

J1: H ALM SET

1:ON

 

8.2.17

High/Low limit alarm value

J2: H ALM VAL

+100.0

%

 

 

 

 

High high/Low low limit alarm On/Off

K1: HH ALM SET

1:ON

 

 

 

 

High high/Low low limit alarm value

K2: HH ALM VAL

+110.0

%

 

 

 

 

8.2.18

Fluid empty alarm

L1: EMPTY ALM

1:NORMAL

 

 

 

 

 

8.2.19

Self-diagnosis On/Off

L2: SELF CHECK

1:ON

 

 

 

 

 

Alarm output factor setting

L3: ALM PRESET

1:WITHOUT EMP

 

 

 

 

 

 

8.2.20

Rate-of-change limit value

M1: LIMIT RATE

05.5

%

Control time

M2: LIMIT TIME

01

s

 

 

 

 

 

 

8.4

Fixed value output (Loop check)

N1: FIXED OUT

OFF

 

N2: FIXED CURR

20.0

mA

 

 

N3: FIXED PULSE

1000

pps

 

 

 

 

 

8.5.2

Zero offset adjustment

O1: MANUAL ZERO

-000.1

%

 

 

 

 

 

8.2Parameters Check / Change

8.2.1Exciting Current Value

The exciting current value is already adjusted when the product is shipped from the factory. Normally the exciting current value is set to the same value as described on the detector nameplate.

Checking the exciting current value

Proceed as follows to check or change the exciting current value.

Switch operation

Display example

Description

 

 

 

 

A1: EX. CURR.

Select A1: EX. CURR. from the setting item selection menu.

 

0.2100A

The currently set exciting current value appears.

A1: EX. CURR.

Returns to the menu of setting item selection.

- 56 -

Page 57
Image 57
Toshiba L5232 manual Parameters Check / Change, Exciting Current Value, Checking the exciting current value