Troubleshooting

To perform the BNC test

To perform the BNC test

Equipment Required

Equipment

Critical Specification

Recommended

 

 

Model/Part

Digitizing Oscilloscope

100 MHz Bandwidth

Agilent 54600A

BNC Shorting Cap

 

1250-0074

BNC Cable

 

Agilent 10503A

BNC-Banana Adapter

 

1251-2277

1Press the Config key.

2Assign pods 1 and 2 to Machine 1.

To assign the pod field, select the pods 1 and 2 field, then select Machine 1 in the pop-up menu.

3In the Analyzer 1 box, select the Type field. Select Timing in the pop-up menu.

4 Set up the trigger menu.

a Press the Trig key. Select Clear Trigger All.

bSelect Arming Control. In the Arming Control pop-up menu, select the field labeled Run, then select Port In. Press the Done key.

5Attach a BNC shorting cap to the External Trigger Input on the rear panel of the logic analyzer.

6Using a BNC cable, connect the External Trigger output to the oscilloscope channel 1 input. Set the oscilloscope to Trigger On and measure TTL voltage levels.

7Press the RUN front panel key.

The warning "MACHINE 1 Waiting on level 1" will appear.

8Remove the shorting cap from the rear panel External Trigger input BNC.

9The warning will go away and the oscilloscope will display a positive-going TTL pulse.

5–27