Testing Performance

To test the single-clock, single-edge, state acquisition

3Set up the oscilloscope.

Oscilloscope Setup

Time Base

Display

Delta V

Delta T

Time/Div: 1.00 ns/div

avg

V markers on

T markers on

 

# of avg: 16

marker 1 position: Chan 1

start on: Pos Edge 1

 

screen: dual

marker 2 position: Chan 1

stop on: Neg Edge 1

 

 

 

 

Channel

 

 

 

 

Channel 1

Channel 2

Display

on

on

Probe Atten

20.00

20.00

Offset

−1.3 V

−1.3 V

Volts/Div

400 mV

400 mV

Set up the logic analyzer

1Set up the Configuration menu. a Press the Config key.

b In the Configuration menu, assign all pods to Machine 1. To assign the pods, select the pod fields, then select Machine 1 in the pop-up menu.

c Select the Type field in the Analyzer 1 box, then select State.

3–24

Page 46
Image 46
Agilent Technologies 01664-97005 manual Set up the oscilloscope, To test the single-clock, single-edge, state acquisition