Testing Performance

To test the single-clock, single-edge, state acquisition

Check the setup/hold combination

1Select the logic analyzer setup/hold time.

a In the logic analyzer Format menu, select Master Clock.

b Select the Setup/Hold field, then select the setup/hold combination to be tested for all pods. The first time through this test, use the top combination in the following table.

Setup/Hold Combinations

3.5/0.0 ns

0.0/3.5 ns

1.5/2.0 ns

cSelect Done to exit the setup/hold combinations.

2Disable the pulse generator channel 2 COMP (with the LED off).

3Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup time of the setup/hold combination selected, +0.0 ps or 100 ps.

a In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker 1 at 1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at 1.3000 V.

b In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Pos Edge 1.

3–29

Page 51
Image 51