Testing Performance

To test the single-clock, multiple-edge, state acquisition

6Test the next clock.

a Press the Format key, then select Master Clock. b Turn off and disconnect the clock just tested.

c Repeat steps 3 and 5 for the next clock listed in the table in step 3, until all listed clock edges have been tested.

7Test the next setup/hold combination. a Press the Format key, then select Master Clock and turn off the clock just tested. b Repeat steps 1 through 7 for the next setup/hold combination listed in step 1, until all listed setup/hold combinations have been tested.

When aligning the data and clock waveforms using the oscilloscope, align the waveforms according to the setup time of the setup/hold combination being tested, +0.0 ps or -100.0 ps.