Testing Performance

To test the glitch capture

3Set up the Trigger menu. a Press the Trigger key.

b Select Modify Trigger, then Clear Trigger, then select All in the pop-up menus.

4Using the Precision Edge Find in the Delta T menu of the oscilloscope, verify that the pulse widths of the pulse generator channels 1 and 2 are 3.450 ns, +50 ps or 100 ps. If necessary, adjust the pulse widths of the pulse generator channels 1 and 2.

5Set up the Waveform menu to view all the channels.

a Select one of the Glitch labels, then select Delete All in the pop-up menu. b Select All, then select continue.

c Press the Select key, then select Insert in the pop-up menu.

d Press the Select key, then select Sequential in the pop-up menu.

3–21

Page 43
Image 43
Agilent Technologies 01664-97005 manual Set up the Waveform menu to view all the channels