3Set up the Trigger menu.
aPress the Trigger key.
bSelect Modify Trigger, then Clear Trigger, then select All in the pop-up menus.
4Using the Precision Edge Find in the Delta T menu of the oscilloscope, verify that
the pulse widths of the pulse generator channels 1 and 2 are 3.450 ns, +50 ps or 100
ps. If necessary, adjust the pulse widths of the pulse generator channels 1 and 2.
5Set up the Waveform menu to view all the channels.
aSelect one of the Glitch labels, then select Delete All in the pop-up menu.
bSelect All, then select continue.
cPress the Select key, then select Insert in the pop-up menu.
dPress the Select key, then select Sequential in the pop-up menu.
Testing Performance
To test the glitch capture
3–21