Testing Performance

To test the time interval accuracy

Acquire the data

1Enable the pulse generator channel 1 output (with the LED off).

2Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic analyzer to acquire data for at least 100 valid runs as indicated in the pattern statistics field.

3When the logic analyzer has acquired at least 100 valid runs, touch Stop. The

Min X-O field in the logic analyzer Pattern Statistics menu should read 94.99–95.00 s. The Max X-O field should read 95.0095.01 s. The Avg X-O field should read 94.99–95.01 s. Record the results in the performance test record.