To test the threshold accuracy

Testing the threshold accuracy verifies the performance of the following specification:

Clock and data channel threshold accuracy.

These instructions include detailed steps for testing the threshold settings of pod 1. After testing pod 1, connect and test pod 2. To test pod 2, follow the detailed steps for pod 1, substituting the pod 2 for pod 1 in the instructions.

Each threshold test tells you to record the voltage reading in the performance test record located at the end of this chapter. To check if each test passed, verify that the voltage reading you record is within the limits listed on the performance test record.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Digital Multimeter

0.1 mV resolution, 0.005% accuracy

Agilent 3458A

Function Generator

DC offset voltage ±6.3 V

Agilent 3325B Option 002

BNC-Banana Cable

 

Agilent 11001-60001

BNC Tee

 

Agilent 1250-0781

BNC Cable

 

Agilent 8120-1840

BNC Test Connector, 17x2

 

 

Set up the equipment

1Turn on the equipment required and the logic analyzer. Let them warm up for 30 minutes before beginning the test.

2Set up the function generator.

a Set up the function generator to provide a DC offset voltage at the Main Signal output.

b Disable any AC voltage to the function generator output, and enable the high voltage output.

c Monitor the function generator DC output voltage with the multimeter.