Testing Performance

To test the threshold accuracy

Set up the logic analyzer

1Press the Config key. Assign all pod fields to Machine 1. To assign the pod fields, select the pod fields, then select Machine 1 in the pop-up menu.

2In the Analyzer 1 box, select the Type field. Select Timing in the pop-up menu.

Connect the logic analyzer

1Using the 17-by-2 test connector, BNC cable, and probe tip assembly, connect the data and clock channels of pod 1 to one side of the BNC Tee.

2Using a BNC-banana cable, connect the voltmeter to the other side of the BNC Tee.

3 Connect the BNC Tee to the Main Signal output of the function generator.

3–9

Page 31
Image 31
Agilent Technologies 01664-97005 Set up the logic analyzer, Connect the logic analyzer, To test the threshold accuracy