To test the time interval accuracy

Testing the time interval accuracy does not check a specification, but does check the following:

125 MHz oscillator

This test verifies that the 125 MHz timing acquisition synchronizing oscillator is operating within limits.

Equipment Required

Equipment

Critical Specifications

Recommended

 

 

Model/Part

Pulse Generator

Function Generator

SMA Cable

Adapter

BNC Test Connector, 6x2

100 MHz 3.5 ns pulse width, < 600 ps rise time Accuracy ≤ (5)(10−6) × frequency

BNC(m)-SMA(f)

Agilent 8131A Option 020 Agilent 3325B Option 002

Agilent 8120-4948

Agilent 1250-2015

Set up the equipment

1Turn on the equipment required and the logic analyzer. Let them warm up for 30 minutes if you have not already done so.

2Set up the pulse generator according to the following table.

Pulse Generator Setup

Channel 1

Period

Mode

EXT TRIG

Delay: 0 ps

5 us

TRIG

Slope: Positive

Width: 2.5 us

 

 

THRE: 1.0 V

High: −0.9 V

 

 

 

Low: −1.7 V

 

 

 

COMP: Disabled

 

 

 

(LED off)

 

 

 