Troubleshooting

To test the logic analyzer probe cables

7Using four 6-by-2 test connectors, four BNC Couplers, and four SMA (m) - BNC (f) Adapters, connect the logic analyzer to the pulse generator channel outputs. To make the test connectors, see chapter 3, "Testing Performance."

a Connect the even-numbered channels of the lower byte of the pod under test to the pulse generator channel 1 Output and J-clock.

b Connect the odd-numbered channels of the lower byte of the pod under test to the pulse generator channel 1 Output.

c Connect the even-numbered channels of the upper byte of the pod under test and the clock channel to the pulse generator channel 2 Output.

dConnect the odd-numbered channels of the upper byte of the pod under test to the pulse generator channel 2 Output.

8On the logic analyzer, press Run. The display should look similar to the figure below.

9If the display looks like the figure, then the cable passed the test.

If the display does not look similar to the figure, then there is a possible problem with the cable or probe tip assembly. Causes for cable test failures include the following:

open channel.

channel shorted to a neighboring channel.

channel shorted to either ground or a supply voltage.

Return to the troubleshooting flowchart.