Troubleshooting

To test the disk drive voltages

W A R N I N G

To test the disk drive voltages

Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and assemblies.

This procedure is to be performed by service-trained personnel aware of the hazards involved, such as fire and electrical shock.

Equipment Required

Equipment

Critical Specification

Recommended

 

 

Model/Part

Digitizing Oscilloscope

> 100 MHz Bandwidth

Agilent 54600A

1Turn off the instrument, then remove the power cable. Remove the instrument cover and the disk drive.

2Reconnect the disk drive cable to the rear of the disk drive. Turn the disk drive over so that the solder connections of the cable socket are accessible.

3Connect the power cable, insert the operating system disk, then turn on the instrument.

4Insert the disk that contains the functional performance verification software and enter the test operating system.

5In the Sys PV menu select the Disk test. Insert a disk that has enough available bytes to run the test in the disk drive, then select Run-Repetitive.

5–25

Page 113
Image 113