Preparing for Use

To degauss the display

1Turn the power switch to the Off position, then remove the power cord from the instrument.

2Remove the fuse module by carefully prying at the top center of the fuse module until you can grasp it and pull it out by hand.

3Reinsert the fuse module with the arrow for the appropriate line voltage aligned with the arrow on the line filter assembly switch.

4Reconnect the power cord. Turn on the instrument by setting the power switch to the On position.

To degauss the display

If the logic analyzer has been subjected to strong magnetic fields, the CRT might become magnetized and display data might become distorted. To correct this condition, degauss the CRT with a conventional external television type degaussing coil.

To clean the logic analyzer

With the instrument turned off and unplugged, use mild soap and water to clean the front and cabinet of the logic analyzer. Harsh soap might damage the water-base paint.

To test the logic analyzer

If you require a test to verify the specifications, start at the beginning of chapter 3, "Testing Performance."

If you require a test to initially accept the operation, perform the self-tests in chapter 3.

If the logic analyzer does not operate correctly, go to the beginning of chapter 5, "Troubleshooting."