Agilent Technologies 01664-97005 manual To test the glitch capture, Set up the pulse generator

Models: 01664-97005

1 164
Download 164 pages 31.78 Kb
Page 39
Image 39

To test the glitch capture

Testing the glitch capture verifies the performance of the following specification:

Minimum detectable glitch.

This test checks the minimum detectable glitch on sixteen data channels at a time.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A Option 020

Digitizing Oscilloscope

≥ 6 GHz bandwidth , < 58 ps rise time

Agilent 54121T

SMA Coax

18 GHz bandwidth

Agilent 8120-4948

(Qty 3)

 

 

Adapter (Qty 4)

SMA(m)-BNC(f)

Agilent 1250-1200

Coupler (Qty 4)

BNC (m)(m)

Agilent 1250-0216

BNC Test Connector, 6x2 (Qty 4)

 

 

Set up the equipment

1Turn on the equipment required and the logic analyzer. Let them warm up for 30 minutes before beginning the test if you have not already done so.

2Set up the pulse generator.

Pulse Generator Setup

Channel 1

Channel 2

Period

Delay: 0 ps

Delay : 0 ps

22.0 ns

Width: 3.5 ns

Width: 3.5 ns

 

High: −0.9 V

High: −0.9 V

 

Low: −1.7 V

Low: −1.7 V

 

3–17

Page 39
Image 39
Agilent Technologies 01664-97005 manual To test the glitch capture, Set up the pulse generator, Pulse Generator Setup