Testing Performance

To test the single-clock, multiple-edge, state acquisition

Connect the logic analyzer

1Using the 6-by-2 test connectors, connect the logic analyzer clock and data channels listed in the following table to the pulse generator. Install a BNC cable between the pulse generator channel 2 output and the 6x2 test connector with the logic analyzer clock leads.

2Using SMA cables, connect channel 1, channel 2, and trigger of the oscilloscope to the pulse generator.

Connect the 1664A Logic Analyzer to the Pulse Generator

Testing

Connect to

Connect to

Connect to

Combination

Agilent 8131A

Agilent 8131A

Agilent 8131A

 

Channel 1 Output

Channel 1

Output

 

Channel 2 Output

1

Pod 1, channel 3

Pod 1, channel 11

J-clock

 

Pod 2, channel 3

Pod 2, channel 11

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3Activate the data channels that are connected according to the previous tables. a Press the Format key.

b Select the field showing the channel assignments for one of the pods being tested. Press the Clear entry key. Using the arrow keys, move the selector to the data channels to be tested, then press the Select key. An asterisk means that a channel is turned on. When all the correct channels of the pod are turned on, press the Done key. Follow this step for the remaining pods.

3–48

Page 70
Image 70
Agilent Technologies 01664-97005 manual Channel