Troubleshooting

To check the power-up tests

To check the power-up tests

The logic analyzer automatically performs power-up tests when you apply power to the instrument (during the boot-up sequence). The revision number of the operating system shows in the upper-right corner of the screen during these power-up tests. As each test completes, either "passed" or "failed" prints on the screen in front of the name of each test.

1Disconnect all inputs, then insert the operating system disk into the disk drive.

2Let the instrument warm up for a few minutes, then cycle power by turning off then turning on the power switch.

If the instrument is not warmed up, the power-up test screen will complete before you can view the screen.

3As the tests complete, check if they pass or fail.

Performing Power-Up Self-Tests

passed

ROM text

passed

RAM test

passed

Interrupt test

passed

Display test

passed

HIL Controller test

passed

HIL Devices test

No Disk

Front Disk test

5–15

Page 103
Image 103
Agilent Technologies 01664-97005 manual To check the power-up tests, As the tests complete, check if they pass or fail