Testing Performance

To test the time interval accuracy

4Set up the Waveform menu. a Press the Waveform key.

b Move the cursor to the sec/Div field, then use the RPG knob to dial in 2.00 s. c Select the Markers Off field, then select Pattern.

d Select the Specify Patterns field. Select X entering 1 and O entering 1.

eSelect Done to exit the Specify Patterns menu.

fMove the cursor to the X-pat field. Type 1, then press Done.

gMove the cursor to the O-pat field. Type 20, then press Done.

hSelect the Markers Patterns field, then select Statistics. Select Reset Statistics to initialize the statistics fields.

Connect the logic analyzer

1Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator channel 1 output.

2Using the SMA cable and the BNC adapter, connect the External Input of the pulse generator to the Main Signal of the function generator.

3–57