Testing Performance

To test the glitch capture

3Set up the oscilloscope.

Oscilloscope Setup

Time Base

Display

Delta V

Delta T

 

Time/Div: 1.00 ns/div

mode: avg

V markers on

T markers on

 

delay: 17.7000 ns

# of avg: 16

marker 1 position: Chan 1

start on: Pos Edge 1

 

 

screens: dual

marker 2 position: Chan 2

stop on: Pos Edge 1

 

 

 

 

 

 

Channel

 

 

 

 

 

Channel 1

Channel 2

Display

on

on

Probe Atten

20.00

20.00

Volts/Div

400 mV

400 mV

Offset

−1.3000 V

−1.3000 V

Set up the logic analyzer

1Press the Config key. Assign all pod fields to Machine 1. To assign the pod fields, select the pod fields, then select Machine 1 in the pop-up menu.

2In the Analyzer 1 box, select the Type field. Select Timing in the pop-up menu.

Connect the logic analyzer

1 Using SMA cables, connect the oscilloscope to the pulse generator channel 1 Output, channel 2 Output, and Trig Output.

2Using the 6-by-2 test connectors, connect the first combination of logic analyzer clock and data channels listed in the table to the pulse generator.

You will repeat this test for the remaining combinations.

3–18

Page 40
Image 40
Agilent Technologies 01664-97005 manual Set up the oscilloscope, To test the glitch capture, Oscilloscope Setup