To adjust the CRT intensity

W A R N I N G

 

Do not touch the CRT monitor sweep board. High voltages exist on the sweep board that can

 

cause personal injury.

 

 

 

 

 

 

 

 

 

 

 

 

 

Equipment Required

 

 

 

 

 

 

 

 

 

Equipment

Critical Specification

Recommended

 

 

 

 

Model/Part

 

 

Alignment Tool

1

8710-1300

 

 

Light Power Meter

 

United Detector 351

1Turn off the logic analyzer, then disconnect the power cord. Remove the cover. Refer to chapter 6, "Replacing Assemblies," for instructions to remove the cover.

2Connect the power cord, insert the operating system disk, then turn on the logic analyzer.

3Access the Display Test.

If you just finished adjusting the CRT monitor alignment, go to step 4.

To Access the Display Test, perform the following steps.

aInsert the disk containing the functional performance verification tests (normally the same disk) into the disk drive, then load the functional performance verification operating system into the logic analyzer (refer to Chapter 3, To Perform The Self Tests).

bEnter the Sys PV tests and enter the Display Test. A grid pattern should appear.

4 Press the front panel Select key.

The display should show a full bright test screen.

5Turn the rear panel intensity adjustment to full bright.

6Place the light power meter against the display at center screen.

4–5

Page 87
Image 87
Agilent Technologies 01664-97005 manual To adjust the CRT intensity, Press the front panel Select key