Testing Performance

To test the single-clock, single-edge, state acquisition

10Using the Delay mode of the pulse generator channel 1, position the pulses according to the setup/hold combination selected, +0.0 ps or 100 ps.

a In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker 1 at 1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at 1.3000 V.

b In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Neg Edge 1.

c Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to the setup time of the setup/hold combination selected.

11Select the clock to be tested.

a In the Master Clock menu, select the clock field to be tested, then select the clock edge as indicated in the table. The first time through this test, use the top clock and edge.

Clocks

J↓

K↓

bConnect the clock to be tested to the pulse generator channel 2 output.

cSelect Done to exit the Master Clock menu (see illustration).

3–32

Page 54
Image 54
Agilent Technologies 01664-97005 manual Select the clock to be tested