Testing Performance

To test the single-clock, single-edge, state acquisition

2Set up the Format menu.

a Press the Format key. Select State Acquisition Mode, then select Full Channel/4K Memory/50MHz.

b Select the field to the right of each pod, then select ECL in the pop-up menu.

3Set up the Trigger menu.

a Press the Trigger key. Select Modify Trigger, then Clear Trigger, then select All in the pop-up menus.

b Select Count Off. Press Select again, then select Time in the pop-up menu. Select Done to exit the menu.

c Select the field labeled 1 under the State Sequence Levels. Select the field labeled "anystate," then select "no state." Select Done to exit the State Sequence Levels menu.

d Select the field next to "a," under the label Lab1. Type "000A", then press the Select key.

3–25

Page 47
Image 47
Agilent Technologies 01664-97005 manual Set up the Format menu, Set up the Trigger menu