Agilent Technologies 01664-97005 manual Performance Test Record

Models: 01664-97005

1 164
Download 164 pages 31.78 Kb
Page 81
Image 81

Performance Test Record

Performance Test Record

 

 

 

 

1664A Logic Analyzer_______

Serial No.______________________

Work Order No.___________________

Recommended Test Interval - 2 Year/4000 hours

Date___________________

Recommended next testing___________________

Temperature___________________

 

 

 

 

 

 

 

 

 

 

 

 

Test

Settings

 

Results

 

 

 

 

 

 

 

Self-Tests

 

 

 

Pass/Fail

________

 

 

 

 

 

 

Threshold

± (100 mV + 3% of

 

 

 

Accuracy

threshold setting)

 

 

 

 

 

 

 

Limits

Measured

Pod 1

TTL, ±145 mV

TTL VL

+1.355 V

________

 

 

 

TTL VH

+1.645 V

________

 

ECL, ±139 mV

ECL VL

-1.439 V

________

 

ECL VH

-1.161 V

________

 

 

 

-User VL

-6.280 V

________

 

-User, ±280 mV

- User VH

-5.720 V

________

 

 

 

+ User VL

+5.720 V

________

 

+User, ±280 mV

+ User VH

+6.280 V

________

 

0 V User VL

-100 mV

________

 

 

 

0 V User VH

+100 mV

________

 

0 V, ±100 mV

 

 

 

Pod 2

TTL, ±145 mV

TTL VL

+1.355 V

________

 

 

 

TTL VH

+1.645 V

________

 

ECL, ±139 mV

ECL VL

-1.439 V

________

 

ECL VH

-1.161 V

________

 

 

 

-User VL

-6.280 V

________

 

-User, ±280 mV

- User VH

-5.720 V

________

 

+ User VL

+5.720 V

________

 

 

 

 

+User, ±280 mV

+ User VH

+6.280 V

________

 

0 V User VL

-100 mV

________

 

 

 

0 V User VH

+100 mV

________

 

0 V, ±100 mV

 

 

 

 

 

 

 

 

 

Glitch Capture

Minimum Detectable

 

 

Pass/Fail

 

Glitch 3.5 ns

 

 

 

 

 

 

Pod 1

 

________

 

 

 

Pod 2

 

________

 

 

 

 

 

 

 

 

 

 

 

 

3–59