Calibrating and Adjusting

To adjust the CRT monitor alignment

4Enter the Sys PV tests, then enter the Display Test.

A grid pattern should appear.

5If the display is tilted (rotated), adjust the CRT yoke by rotating it to straighten the display.

Yoke

6If the grid pattern is not centered horizontally, adjust the H-Hold.

7 If you need to adjust the intensity, go to the next page.

If you are finished with the adjustments, turn off the instrument, then remove the power cord. Install the cover on the instrument.